MPMS SQUID Magnetometer
This is a guide on how to center the sample in MPMS SQUID Magnetometer with RSO mode before your actual measurement.
Click is when you need to click the bottom on the Computer.
-
Sample Removal:
On computer: top bar Sample -> Remove
On SQUID:
- Lower the black knob, twist, such that it connects to the black wire that is attached to the sample plate.
- Pull the knob + wire up.
- Close sample chamber (turn the black knob on the chamber to close)
- Pull sample tube out, and swap with the cap (make sure SQUID will not be contaminated).
- Swap samples on the sample plate.
Optional on computer: Sample -> Remove vent
-
Sample loading:
On SQUID:
- Swap sample tube with the black cap.
- Purge 3 times.
- Open sample chamber, and purge again.
- Lower the sample, turn the black knob to release the sample.
On computer: Sample -> Finished.
Input sample descriptions.
-
Centering: (for small samples)
Set field: set field to 10,000 -> no overshoot, HiRes
RSO centering:
- Parameter -> 6cm
- Initialize transport
- Center
- Adjust position -> auto adjust (want center to be 3cm)
- Parameter -> 3cm
- Adjust position -> auto adjust (until the center is at 1.5cm, at least < 1.55cm)
- Close
If not successful, lower temperature to base, and try again.
Want to use 3cm for RSO scans, want to use 6cm for DC scans.
-
Set Temperature
- Run sequence: any Degauss sequence.
- Set 100K with -5K/min.
- Lower to your target temperature, and start your actual scan.